Development of a Device Equipped with a Thermocouple for Measuring Nasal Airflow
Seung Kyu Chung, Seok Joo Koh, Hyun Joon Lim
1Department of Otorhinolaryngology, Sungkyunkwan University School of Medicine, Samsung Medical Center, Seoul, Korea. skchung@smc.samsung.co.kr
2Department of Otorhinolaryngology, Hanil General Hospital,Seoul, Korea.
3Department of Otorhinolaryngology, College of Medicine, Hallym University, Seoul, Korea.
Abstract
Background and Objectives
To measure nasal cavity function as an airway, rhinomanometry and acoustic rhinometry are currently being used in clinical settings. However, these methods are not helpful for continuously measuring the aerodynamic status of both nasal cavities simultaneously. Therefore, a new instrument to evaluate the nasal flow is required.
MATERIALS AND METHODS
To measure the airflow of bilateral nasal cavity simultaneously, two thermocouples are held in the headset, with the tips of thermocouples positioned below nostrils. The thermocouples are connected to the analog-digital converter and the digitized data is transferred to a notebook computer, in which a graphical programming language software is installed. Eighteen adults were recruited for this study who had no structural abnormality in their nasal cavities. For every subject, measurements from acoustic rhinometry, the thermocouple device, and rhinomanometry were taken in succession. The data from the thermocouple device was compared with those taken from the acoustic rhinometry and rhinomanometry.
Results
A negative correlation was noted between the minimum cross-sectional area by acoustic rhinometry and the inspiratory slope by thermocouple. No correlation was noted between the results for rhinomanometry and the thermocouple device.
Conclusion
The thermocouple device has some advantages over other devices for its non-invasive, continuous, and real-time measurements and its ability to measure bilateral nasal cavity simultaneously.
Key Words: Thermocouple;Nasal flow;Temperature;Acoustic rhinometry;Rhinomanometry


TOOLS
METRICS Graph View
  • 838 View
  • 0 Download
Related articles


Editorial Office
101 Hyundai ESA Apt., 20, Hyoryeong-ro 77-gil, Seocho-gu, Seoul 06628, Republic of Korea
Tel: +82-2-3461-9945    Fax: +82-2-3461-9947    E-mail: office@j-rhinology.org                

Copyright © 2024 by Korean Rhinologic Society.

Developed in M2PI

Close layer
prev next